• Model Designation. TR9292 (Aspiration Duct Probe) with the "-A" option specifying simultaneous 0–5V and 4–20mA linear analog outputs and no LCD display.
• Sensing Method. Gold-plated Non-Dispersive Infrared (NDIR) optical sensor with Automatic Baseline Correction (ABC) self-calibration.
• Measurement Range. 0 to 2,000 ppm CO2 (factory adjustable up to 10,000 ppm upon request).
• Accuracy. ±1% of measurement range plus ±3% of measured value.
• Response Time. T90 < 15 seconds (flow-through) / < 2 minutes (diffusion).
• Output Signals. Simultaneous dual analog outputs: 0–5 VDC and 4–20 mA (requires common ground with the control system).
• Power Supply. 24 VAC/VDC ±20%, 50–60 Hz (half-wave rectified), drawing ≤ 1 Watt average.
• Operating Environment. 32°F to 122°F (0°C to 50°C); 0 to 95% non-condensing relative humidity.
• Storage Environment. -40°F to 158°F (-40°C to 70°C).
• Rated Life. Minimum 15 years without requiring manual calibration.
-